Automated titanium fastener for surgical aortic valve replacement—preventive role for infective endocarditis? (2024)

  • Journal List
  • Eur J Cardiothorac Surg
  • PMC11211209

As a library, NLM provides access to scientific literature. Inclusion in an NLM database does not imply endorsem*nt of, or agreement with, the contents by NLM or the National Institutes of Health.
Learn more: PMC Disclaimer | PMC Copyright Notice

Automated titanium fastener for surgical aortic valve replacement—preventive role for infective endocarditis? (1)

Link to Publisher's site

Eur J Cardiothorac Surg. 2024 Jun; 65(6): ezae236.

Published online 2024 Jun 24. doi:10.1093/ejcts/ezae236

PMCID: PMC11211209

PMID: 38913864

Amila Kahrovic,Automated titanium fastener for surgical aortic valve replacement—preventive role for infective endocarditis? (2) Harald Herkner, Philipp Angleitner, Paul Werner, Alfred Kocher, Marek Ehrlich, Dominik Wiedemann, Guenther Laufer, Paul Simon, and Martin Andreas

Author information Article notes Copyright and License information PMC Disclaimer

Associated Data

Supplementary Materials
Data Availability Statement

Abstract

OBJECTIVES

Evidence on long-term clinical outcomes considering suture-securing techniques used for surgical aortic valve replacement is still uncertain.

METHODS

A total of 1405 patients who underwent surgical aortic valve replacement between January 2016 and December 2022 were included and grouped according to the suture-securing technique used (automated titanium fastener versus hand-tied knots). The occurrence of infective endocarditis during follow-up was set as the primary study end-point. As secondary study end-points, stroke, all-cause mortality and a composite outcome of either infective endocarditis, stroke, or all-cause mortality were assessed.

RESULTS

The automated titanium fastener was used in 829 (59%) patients, whereas the hand-knot tying technique was used in 576 (41%) patients. The multivariable proportional competing risk regression analysis showed a significantly lower risk of infective endocarditis during follow-up in the automated titanium fastener group (adjusted sub-hazard ratio 0.44, 95% confidence interval 0.20–0.94, P = 0.035). The automated titanium fastener group was not associated with an increased risk of mortality or attaining the composite outcome, respectively (adjusted hazard ratio 0.81, 95% confidence interval 0.60–1.09, P = 0.169; adjusted hazard ratio 0.82, 95% confidence interval 0.63–1.07, P = 0.152). This group was not associated with an increased risk of stroke (adjusted sub-hazard ratio 0.82, 95% confidence interval 0.47–1.45, P = 0.504). Also, a significantly lower rate of early-onset infective endocarditis was observed in the automated titanium fastener group, (0.4% vs 1.4%, P = 0.032).

CONCLUSIONS

Suture-securing with an automated titanium fastener device appears to be superior compared to the hand-knot tying technique in terms of lower risk of infective endocarditis.

Keywords: Automated titanium fastener, Hand-tied knots, Surgical aortic valve replacement, Infective endocarditis

Surgical aortic valve replacement (SAVR) is a routinely performed procedure to treat different pathologies of the aortic valve, including aortic valve stenosis, aortic valve regurgitation or a combination of both stenosis and regurgitation.

Graphical abstract

Open in a separate window

INTRODUCTION

Surgical aortic valve replacement (SAVR) is a routinely performed procedure to treat different pathologies of the aortic valve, including aortic valve stenosis, aortic valve regurgitation or a combination of both stenosis and regurgitation. When indicated, SAVR can be performed with either biological or mechanical valve prosthesis [1].

In the context of aortic valve prosthesis fixation, sutures can be secured using an automated titanium fastener device (Cor-Knot Device, LSI Solutions, Victor, New York, USA) or a conventional technique of hand-knot tying. The automated titanium fastener device is designed to facilitate suture-securing through pivotal steps including the advancement of the device over slightly tensioned suture to the designated position, then the lever of the device is squeezed, releasing the crimped automated titanium fastener, and simultaneously trimming excess suture tails [2].

The objectives of this study were to comprehensively analyse the long-term clinical outcomes of patients who underwent SAVR with respect to the suture-securing technique (automated titanium fastener versus hand-tied knots).

PATIENTS AND METHODS

Ethical statement

The Ethics Committee of the Medical University Vienna approved this study, in compliance with the Declaration of Helsinki; (vote number 1243/2023; approval date: 8 May 2023). Informed consent was waived.

Study design

A total of 2465 consecutive adult patients who underwent SAVR between January 2016 and December 2022 at the Department of Cardiac Surgery, Medical University of Vienna were screened for inclusion and exclusion criteria. The exclusion criteria were surgery of mitral, tricuspid or pulmonary valve in combination with SAVR, Bentall procedure, replacement of ascending aorta, previous cardiac surgery and infective endocarditis (IE) at baseline; Supplementary Material, Fig. S1. One thousand four hundred five patients were considered as eligible for inclusion in this retrospective study and grouped according to the suture-securing technique used (automated titanium fastener versus hand-tied knots) during SAVR. Identification of suture-securing technique relied on a comprehensive review of operation reports and institutional surgery protocols. The decision to use either suture-securing technique was driven by the individual preferences of the operating surgeon.

Study endpoints and definitions

The occurrence of IE during follow-up was set as the primary study endpoint. As secondary endpoints, the occurrence of stroke, all-cause mortality and a composite outcome (defined as the occurrence of either IE, stroke or all-cause mortality) were assessed. The diagnosis of IE was defined following the modified Duke criteria [3], irrespective of whether surgical or conservative treatment was pursued. The occurrence of IE of the aortic valve prosthesis was categorized as early-onset, if the infection developed within the initial 12 months of SAVR and as late-onset, if it occurred after 1 year [3]. Stroke was defined as the manifestation of a prolonged or permanent neurological impairment, confirmed by the acute lesion on a neuroimaging scan [4]. All-cause mortality encompassed all deaths following SAVR, regardless of the underlying cause. Survival data were retrieved from federal statistics (Statistics Austria). Study endpoints were extracted from patients’ medical records. A complete dataset was generated using a prospectively managed database at our institution. A total of 4.6% (n = 64) patients were considered as lost to follow-up. The follow-up period was terminated upon the occurrence of each study endpoint, the death of the patient, the last date of hospital visit for those categorized as lost to follow-up, or upon the end of the study. The last date of follow-up was 28 September 2023.

Statistical analysis

Data are presented as median and interquartile range (IQR) for continuous variables, and frequencies and percentages for categorical variables. Statistical calculations comparing continuous variables were made with the Mann–Whitney U-test for not normally distributed variables. Comparisons of categorical variables were made using the chi-square test. To evaluate the crude and adjusted effect of the suture-securing techniques (automated titanium fastener versus hand-tied knots) on the study endpoints, univariable and multivariable proportional hazards regression analyses were performed. The exposure time started from the date of the surgery. The selection of co-variables for the multivariable models was based on a causal concept, considering published evidence and clinical experience.

The multivariable proportional subhazards Fine and Gray regression model accounting for death as a competing event was used to analyse the effect of the suture-securing technique on IE during follow-up. The estimated subhazard ratios (sHR) with 95% confidence intervals (CI) were provided. The following co-variables were included in the model: age (years2), body mass index (kg/m2), dialysis, bicuspid aortic valve, urgent surgery, full sternotomy, biological valve prosthesis and coronary artery bypass grafting. As a sensitivity analysis, the influence of suture-securing technique type on the risk of IE was assessed using inverse-probability-weighted regression adjustment. However, the results of this model were in agreement with those from multivariable proportional competing risk regression model and therefore not reported.

To quantify the model performance, we calculated Harrel’s C, where possible. Based on statistical metrics, the multivariable regression model was deemed appropriate. Kaplan–Meier cumulative event curves were generated to visualize the probability of IE between the study groups. Additionally, we tested for an effect modification of the main effect by different aortic valve prostheses (valve-specific sub-groups) by including an interaction term into the models.

For the secondary study endpoint, stroke, the multivariable proportional subhazards Fine and Gray regression analysis accounting for death as a competing event was used. We selected the following co-variables in the model: era (2016–2019 years), age (years2), history of stroke, cerebrovascular disease, atrial fibrillation, dialysis and biological valve prosthesis.

Furthermore, the multivariable Cox proportional hazards regression models were generated to investigate the effects of the suture-securing technique on all-cause mortality and composite outcome. The effects were reported as an adjusted hazard ratio (HR) with 95% CI. In the analysis of all-cause mortality, the model was adjusted for European System for Cardiac Operative Risk Evaluation II (EuroSCORE II) (log-transformed), era (2016–2019 years), body mass index (kg/m2), biological valve prosthesis and cardiopulmonary bypass time (min). For the assessment of the composite outcome, the model included the following co-variables: era (2016–2019 years), age (years2), EuroSCORE II (log-transformed), diabetes mellitus, dialysis, cerebrovascular disease and biological valve prosthesis.

The statistical analysis was performed using SPSS software 27.0 (IBM Corp, Armonk, NY, USA) and STATA 16.1 software (StataCorp LLC, College Station, TX, USA). A two-sided P-value of <0.05 was considered as statistically significant.

RESULTS

Baseline characteristics

The automated titanium fastener was used in 829 patients (59%), while the hand-knot tying technique was performed in 576 patients (41%). Patients in the automated titanium group were younger (median 70.5, IQR 62.4–76.2 versus median 72.1, IQR 64.1–77.5, P = 0.007) and were presented with a higher body mass index (median 28.0, IQR 25.3–31.7 versus median 27.5, IQR 24.8–30.9, P = 0.034; Table1). All other baseline characteristics variables were comparable between study groups. Preoperative echocardiography data are presented in the Supplementary Material, Table S5. In terms of aortic valve anatomy, the automated titanium fastener group was presented with a lower rate of a tricuspid aortic valve (65.9% vs 76.0%, P < 0.001), and a higher rate of a bicuspid aortic valve (32.6% vs 22.7%, P < 0.001).

Table 1:

Baseline characteristics

VariablesAutomated titanium fastener, N = 829 (59%)Hand-tied knots, N = 576 (41%)P-value
Age (years) (25th–75th interval)70.5 (62.4–76.2)72.1 (64.1–77.5)0.007
Female (%)271 (32.7)211 (36.6)0.126
Body mass index (kg/m2) (25th–75th interval)28.0 (25.3–31.7)27.5 (24.8–30.9)0.034
EuroSCORE II, 25th–75th interval1.8 (1.1–3.4)2.1 (1.1–3.6)0.054
NYHA class III–IV, n (%)300 (36.2)230 (39.9)0.155
Hypertension, n (%)726 (87.6)489 (84.9)0.149
Atrial fibrillation, n (%)136 (16.4)96 (16.7)0.897
 Paroxysmal102 (12.3)69 (12.0)0.855
 Persistent34 (4.1)27 (4.7)0.596
Previous pacemaker implantation, n (%)25 (3.0)19 (3.3)0.765
Diabetes mellitus, n (%)232 (28.0)160 (27.8)0.932
Peripheral vascular disease, n (%)84 (10.1)46 (8.0)0.172
Cerebrovascular disease, n (%)161 (19.4)93 (16.1)0.117
History of stroke, n (%)42 (5.1)24 (4.2)0.433
Dialysis, n (%)6 (0.7)8 (1.4)0.217

Open in a separate window

Bold indicates statistical significance (P < 0.05).

EuroSCORE II: European System for Cardiac Operative Risk Evaluation II; NYHA: New York Heart Association.

Operative characteristics

In the 1st study era (2016–2019), the hand-knot tying technique was used more frequently (75.9%) compared to automated titanium fastener (51.5%). However, in the subsequent era (2020–2022), there was a significant shift towards increased use of automated titanium fastener, (48.6% vs 24.1%, P ≤ 0.001; Table2). The automated titanium fastener group underwent urgent SAVR at a significantly lower rate (10.5% vs 14.1%, P = 0.043). Concerning surgical access, this group had a lower rate of full-sternotomy (47.4% vs 59.7%, P < 0.001) and a higher rate of hemi-sternotomy and thoracotomy, respectively (36.8% vs 31.4%, P = 0.037; 15.8% vs 8.9%, P ≤ 0.001). No differences were found among the study groups regarding the use of mechanical or biological prostheses. As a concomitant procedure, coronary artery bypass grafting was performed at a lower rate in the automated titanium fastener group (31.7% vs 41.0%, P ≤ 0.001). Details on implanted aortic valve prostheses are provided in Supplementary Material, Table S6.

Table 2:

Operative characteristics

VariablesAutomated titanium fastener, N = 829 (59%)Hand-tied knots, N = 576 (41%)P-value
Era, n (%)<0.001
 2016–2019426 (51.5)437 (75.9)
 2020–2022403 (48.6)139 (24.1)
Urgency status, n (%)
 Elective729 (87.9)487 (84.5)0.067
 Urgent87 (10.5)81 (14.1)0.043
 Emergency13 (1.6)8 (1.4)0.785
 Salvage0 (0.0)0 (0.0)
Access, n (%)
 Full-sternotomy393 (47.4)344 (59.7)<0.001
 Hemi-sternotomy305 (36.8)181 (31.4)0.037
 Thoracotomy131 (15.8)51 (8.9)<0.001
Aortic valve replacement, n (%)0.830
 Mechanical prosthesis101 (12.2)68 (11.8)
 Biological prosthesis728 (87.8)508 (88.2)
Concomitant procedures, n (%)
 CABG263 (31.7)236 (41.0)<0.001
 Aortic reduction plasty81 (9.8)62 (10.8)0.545
 Aortic root enlargement27 (3.3)24 (4.2)0.370
 Septal myectomy46 (5.5)34 (5,9)0.778
 Atrial fibrillation surgery46 (5.5)36 (6.3)0.581
 Left atrial appendage resection58 (7.0)54 (9.4)0.105
CPB time (min), 25th–75th interval116 (95–148)115 (92–149)0.134
Total operating room time (min), 25th–75th interval245 (203–298)250 (203–300)0.613

Open in a separate window

Bold indicates statistical significance (P < 0.05).

CABG: Coronary artery bypass grafting; CPB: cardiopulmonary bypass.

Postoperative adverse events

No differences were found among the study groups, regarding postoperative adverse events; Supplementary Material, Table S7.

Long-term outcomes

The rate of ischaemic and haemorrhagic stroke during follow-up was similar between the study groups (Table3). The rate of reintervention was lower in the automated titanium fastener group (1.3% vs 3.5%, P = 0.007). Indications for reintervention included paravalvular leak, structural valve deterioration and IE. No statistically significant difference was observed in the rate of reoperation due to paravalvular leak between the automated titanium fastener group and the hand-tied knots group (0.4% vs 1.0%, P = 0.116). Irrespective of the chosen therapeutic approach (surgery or no surgery), IE of the aortic valve prosthesis occurred less frequently in the automated titanium fastener group (1.1% vs 3.0%, P = 0.011). Also, this group had a significantly lower rate of early-onset IE (0.4% vs 1.4%, P = 0.032). Concerning other reoperations or transcatheter procedures during follow-up, no differences were found among the study groups (Supplementary Material, Table S8).

Table 3:

Long-term outcomes

VariablesAutomated titanium fastener, N = 829 (59%)Hand-tied knots, N = 576 (41%)P-value
Stroke
 Ischaemic, n (%)22 (2.7)21 (3.6)0.288
 Haemorrhagic, n (%)5 (0.6)3 (0.5)0.840
Reintervention, n (%)11 (1.3)20 (3.5)0.007
 Paravalvular leak3 (0.4)6 (1.0)0.116
 Structural valve deterioration3 (0.4)5 (0.9)0.215
 Valve thrombosis0 (0.0)0 (0.0)
 Infective endocarditis (surgery)5 (0.6)9 (1.6)0.075
Infective endocarditis (no surgery), n (%)4 (0.5)8 (1.4)0.069
Infective endocarditis (all cases), n (%)9 (1.1)17 (3.0)0.011
 Early-onset3 (0.4)8 (1.4)0.032
 Late-onset6 (0.7)9 (1.6)0.132

Open in a separate window

Bold indicates statistical significance (P < 0.05).

Study endpoints

Primary study endpoint

The risk of IE during follow-up was significantly lower in the automated titanium fastener group, as indicated in the univariable proportional competing risk regression analysis (sHR 0.42, 95% CI 0.19–0.93, P = 0.033; Table4). Additionally, the multivariable proportional competing risk regression analysis demonstrated a significantly lower risk of IE in this group (adjusted sHR 0.44, 95% CI 0.20–0.94, P = 0.035; Table4; Supplementary Material, Table S1). The primary study endpoint was visually represented by the Kaplan–Meier cumulative event curves of both study groups (Fig.1). The effect was not modified by different aortic valve prostheses (valve-specific sub-groups) in any of the interaction analyses. The median follow-up duration was 1239 days (IQR 706–1829) for the automated titanium fastener group and 1686 days (IQR 842–2309) for the hand-tied knots group.

Open in a separate window

Figure 1:

Kaplan–Meier cumulative event curves show the probability of IE between the automated titanium fastener group versus hand-tied knots group. CI: confidence interval; IE: infective endocarditis.

Table 4:

Study endpoints

Automated titanium fastener, N = 829 (59%)Hand-tied knots, N = 576 (41%)Univariable relative effectsMultivariable relative effects
95% CIP-value95% CIP-value
Infective endocarditisa9 (1.1)17 (3.0)0.42 (0.19–0.93)0.0330.44 (0.20–0.94)0.035
Strokea27 (3.3)24 (4.2)0.84 (0.48–1.45)0.5240.82 (0.47–1.45)0.504
All-cause mortalityb86 (10.4)100 (17.4)0.76 (0.57–1.01)0.0630.81 (0.60–1.09)0.169
Composite outcomeb109 (13.1)124 (21.5)0.74 (0.57–0.96)0.0240.82 (0.63–1.07)0.152

Open in a separate window

Bold indicates statistical significance (P < 0.05).

aEffects calculated as sHR based on multivariable proportional competing risk regression model.

bEffects calculated as HR based on multivariable Cox proportional hazards regression model.

CI: confidence interval; HR: hazard ratio; sHR: sub-hazard ratio.

Secondary study endpoints

The use of an automated titanium fastener was not associated with an increased risk of stroke during follow-up, as evidenced in the multivariable proportional competing risk regression analysis (adjusted sHR 0.82, 95% CI 0.47–1.45, P = 0.504; Table4; Supplementary Material, Table S2). Notably, older age and cerebrovascular disease at baseline were independently associated with an increased risk of stroke, respectively (adjusted sHR 1.00, 95% CI 1.00–1.01, P = 0.009; adjusted sHR 2.04, 95% CI 1.10–3.79, P = 0.024; Supplementary Material, Table S2).

Overall, 186 patients (13.2%) died during follow-up. The multivariable Cox proportional hazards regression model showed that the use of automated titanium fastener was not associated with an increased risk of all-cause mortality (adjusted HR 0.81, 95% CI 0.60–1.09, P = 0.169) (Table4; Supplementary Material, Table S3).

The risk of experiencing the composite outcome was significantly lower in the automated titanium fastener group based on the univariable Cox proportional hazards regression analysis (adjusted HR 0.74, 95% CI 0.57–0.96, P = 0.024; Table4). Further, in the multivariable Cox proportional hazards regression model, the automated titanium fastener group was not associated with an increased risk of attaining the composite outcome during follow-up (adjusted HR 0.82, 95% CI 0.63–1.07, P = 0.152; Table4; Supplementary Material, Table S4).

Clinical presentation of patients presenting with infective endocarditis during the follow-up

A total of 26 patients (1.9%), identified during the follow-up period, met the criteria for definite IE diagnosis, in accordance with the modified Duke criteria and clinical guidelines [3] (Supplementary Material, Figs S2 and S3). Prevalence rates of major and minor criteria are presented in Supplementary Material, Table S9. Furthermore, Supplementary Material, Table S10 lists the pathogens detected in the blood culture results and intraoperative smears (if applicable—patients who underwent surgery for IE).

Causes of death

Cardiac-related mortality emerged as the predominant cause of death, with respiratory causes and multiorgan failure following in succession (Table5).

Table 5:

Causes of death

VariablesAutomated titanium fastener, N = 829 (59%)Hand-tied knots, N = 576 (41%)P-value
Cardiac, n (%)34 (4.1)31 (5.4)0.261
Respiratory, n (%)13 (1.6)18 (3.1)0.051
 COVID-196 (0.7)10 (1.7)0.079
Multi-organ failure, n (%)14 (1.7)13 (2.3)0.445
Infection, n (%)5 (0.6)8 (1.4)0.130
Cancer, n (%)7 (0.8)8 (1.4)0.329
Neurological, n (%)3 (0.4)7 (1.2)0.061
Other, n (%)4 (0.5)5 (0.9)0.373
Unknown, n (%)6 (0.7)10 (1.7)0.079

Open in a separate window

COVID-19: Coronavirus disease 2019.

DISCUSSION

This study represents the 1st comparative analysis of long-term clinical outcomes among a cohort of 1405 patients, with a focus on the suture-securing technique used (automated titanium fastener versus hand-tied knots) during SAVR. The major finding of the study is that the use of an automated titanium fastener device was associated with a lower risk of IE during follow-up (sHR 0.44, 95% CI 0.20–0.94, P = 0.035).

One of the primary advantages of employing an automated titanium fastener device might be the potential reduction in contamination through the implementation of the no-touch principle. Importantly, the suture-securing with an automated titanium fastener device is achieved through a single deployment, with no direct hand-contact with the newly implanted valve prosthesis (Fig.2a). In contrast, the hand-knot tying technique involves multiple, direct hand-touches for securing 1 suture, resulting in over a hundred direct hand-touches on the newly implanted valve prosthesis (Fig.2b). Following extended surgical procedure, it is pertinent to note that surgical gloves might bear pathogens and subsequently, disseminate these while tying knots manually [5]. Surgical site contamination, including the newly implanted prosthesis, might lead to postoperative occurrence of early-onset IE [6]. Indeed, among the pathogens detected in the blood culture results of patients presenting with IE during follow-up (Supplementary Material, Table S10), several belong to the normal flora of the skin, including Staphylococci and Streptococci species [7].

Open in a separate window

Figure 2:

Suture-securing techniques during SAVR (automated titanium fastener versus hand-tied knots). (a) Deployment of an automatic titanium fastener with no direct hand-contact on the newly implanted aortic valve prosthesis. The arrow indicates the position of the automatic titanium fastener device at the target site. (b) Tying of the knot manually, with the focus on direct hand-contact on the newly implanted aortic valve prosthesis. (c) Enlarged view of the flat, smooth, uniform surface of automated titanium fastener crimped around the suture. (d) Enlarged representation of hand-tied knots with numerous cervices and throws for potential bacterial attachment. The total number of throws amounts to 8. SAVR: surgical aortic valve replacement.

Secondary, surface characteristics should not be underestimated, in the context of bacterial attachment [8, 9]. Irregularities in topography, as observed in braided sutures with multiple knots, introduce numerous microenvironments where bacteria can adhere and thrive [10, 11] (Fig.2d). Also, multiple tied, braided sutures may provide an increased surface area for bacterial attachment compared to the smoother, flatter surface of an automated titanium fastener [11, 12]. The establishment of the bacterial attachment, either due to contamination or during transient bacteraemia, is an initial stage in the development of an infection [13]. Following bacterial adherence to the surface, bacteria tend to multiply in a more organized and cooperative manner within an extracellular polysaccharide slime-like matrix, forming a structured biofilm. The biofilm creates a protective environment that allows bacteria to grow, impede antibiotic efficacy and the host’s immune defences, as well as exhibit increased virulence [8, 13]. Conversely, the flat, uniform surface of the automated titanium fastener, crimped around the suture might be less conducive to bacterial adherence (Fig.2c).

Additionally, a significantly lower rate of early-onset IE of the aortic valve prosthesis was observed in the automated titanium fastener group (0.4% vs 1.4%, P = 0.032). Concerning the pathogenesis of early-onset IE, in instances of contamination, infection usually involves braided sutures and the sewing ring, expanding to the annulus, and resulting in the formation of IE-related lesions (abscesses, pseudoaneurysms and fistulae), as stated in the 2023 European Society of Cardiology Guidelines for the Management of Infective Endocarditis [3].

The current analysis indicates that the automated titanium fastener was not associated with an increased risk of stroke during follow-up (adjusted sHR 0.82, 95% CI 0.47–1.45, P = 0.504). Several studies reported no difference in the rate of short-term stroke onset among the study groups [14, 15]. Indeed, surgical titanium is a commonly used material in medical devices, primarily due to its favourable mechanical properties and biocompatibility. Although surgical titanium may have thrombogenic potential, the subsequent neo-endothelialization of the surface might enhance its haemocompatibility, thus mitigating the risk of blood clot formation or other adverse events, such as haemolysis [16, 17]. Furthermore, Li et al. conducted an experimental study using an ovine model to assess the biocompatibility, local tissue pathological changes, inflammatory response and thrombosis tendency of an automated titanium fastener compared to hand-tied knots. The study found that neo-endothelialization was comparable among the study groups by day 60. Notably, no thrombus formation was detected on the surface of the automated titanium fastener [18].

Concerning mortality, the systematic review and meta-analysis examined mortality between the automated titanium fastener group and the hand-tied knots group, indicating no significant difference [19], which is consistent with our findings.

Suture-securing with an automated titanium fastener provides uniform tension along the prosthesis [2]. An experimental study by Lee et al. demonstrated that sutures secured with automated titanium fastener are significantly more consistent, on average twice as strong as hand-tied knots, which might reduce the risk of paravalvular leak formation [4]. Indeed, the occurrence of the paravalvular leak might have a technical nature and might be attributed to the undesirable loose hand-tied knots or ‘air-knots’. In the present analysis, the rate of reoperation due to the paravalvular leak was lower in the automated titanium fastener group; however, the significance level has not been reached (0.4% vs 1.0%, P = 0.116).

The cases of leaflet perforation attributed to automated titanium fastener have been reported in a few case reports [20]. To ensure safe deployment of the automated titanium fastener, the suture slot and the rotational knob’s indicator fin should be consistently aligned towards the centre of the prosthesis. Noteworthy, in the present study involving 829 patients in the automated titanium fastener group, no such complication was documented.

In terms of economics, the cost of the automated titanium fastener device is ∼€165 for the European market, with an additional charge of ∼€45 for each individual loading unit used for the fixation of a single stitch. The financial implication of implementing such technology is essential for healthcare institutions as it may infuence decision-making processes.

Limitations

The present analysis is a single-centre retrospective study. Only patients who underwent isolated SAVR were included in the study, which may not represent the real patient population. Given the retrospective nature of the study, notable selection bias is apparent, manifested in the disparity in age, era, aortic valve anatomy, surgical access, urgent setting and concomitant coronary artery bypass grafting surgery. The use of either suture-securing technique was surgeon-dependent. The number of stitches utilized for prosthesis fixation was variable though different types and sizes of aortic valve prostheses. Information regarding minor paravalvular leak, as the possible predisposition for IE, was not available for statistical analysis. The occurrence of IE during follow-up can be caused by several factors, independent of the suture-securing technique used. Only patients with a definite diagnosis of IE (following modified Duke criteria and current guidelines [3]) were considered for the analysis. The current literature on this topic is limited.

CONCLUSIONS

The automated titanium fastener device seems to be a promising surgical tool not only for facilitating SAVR, but also potentially reducing the risk of IE compared to the hand-knots tying technique. Avoiding direct hand-contact on the valve prosthesis by implementing the ‘no-touch principle’ with an automated titanium fastener device in valve surgery might be beneficial. Also, the importance of suture characteristics for potential bacterial adherence should not be underestimated. A prospective randomized controlled trial is warranted to confirm our findings.

Supplementary Material

ezae236_Supplementary_Data

Click here to view.(14M, docx)

Glossary

ABBREVIATIONS

CIConfidence interval
EuroSCORE IIEuropean System for Cardiac Operative Risk Evaluation II
HRHazard ratio
IEInfective endocarditis
IQRInterquartile range
SAVRSurgical aortic valve replacement
sHRSubhazard ratio

Contributor Information

Amila Kahrovic, Department of Cardiac Surgery, Medical University of Vienna, Vienna, Austria.

Harald Herkner, Department of Emergency Medicine, Medical University of Vienna, Vienna, Austria.

Philipp Angleitner, Department of Cardiac Surgery, Medical University of Vienna, Vienna, Austria.

Paul Werner, Department of Cardiac Surgery, Medical University of Vienna, Vienna, Austria.

Alfred Kocher, Department of Cardiac Surgery, Medical University of Vienna, Vienna, Austria.

Marek Ehrlich, Department of Cardiac Surgery, Medical University of Vienna, Vienna, Austria.

Dominik Wiedemann, Department of Cardiac Surgery, Medical University of Vienna, Vienna, Austria.

Guenther Laufer, Department of Cardiac Surgery, Medical University of Vienna, Vienna, Austria.

Paul Simon, Department of Cardiac Surgery, Medical University of Vienna, Vienna, Austria.

Martin Andreas, Department of Cardiac Surgery, Medical University of Vienna, Vienna, Austria.

SUPPLEMENTARY MATERIAL

Supplementary material is available at EJCTS online.

FUNDING

This study received no funding.

Conflict of interest: Dr Martin Andreas has received institutional research funding (Edwards, Abbott, LSI, Medtronic) and has served as a proctor/speaker/consultant (Edwards, Abbott, Medtronic, Zoll, Boston, Abbvie). Dr Dominik Wiedemann has served as a proctor/consultant for Abbott and Scientific Advisor for Xenios.

DATA AVAILABILITY

Anonymized datasets will be shared on reasonable request to the corresponding author.

Author contributions

Amila Kahrovic: Conceptualization; Data curation; Investigation; Writing—original draft. Harald Herkner: Conceptualization; Formal analysis; Methodology; Writing—original draft. Philipp Angleitner: Conceptualization; Writing—original draft. Paul Werner: Conceptualization; Writing—original draft. Alfred Kocher: Supervision; Writing—review and editing. Marek Ehrlich: Supervision; Writing—review & editing. Dominik Wiedemann: Supervision; Writing—review and editing. Guenther Laufer: Supervision; Writing—review & editing. Paul Simon: Resources; Supervision; Writing—review & editing. Martin Andreas: Conceptualization; Supervision; Writing—review & editing.

Reviewer information

European Journal of Cardio-Thoracic Surgery thanks W. Brent Keeling, Ulrich Otto von Oppell and the other anonymous reviewers for their contribution to the peer review process of this article.

The abstract has been submitted to the 38th EACTS Annual Meeting.

REFERENCES

1. Vahanian A, Beyersdorf F, Praz F, Milojevic M, Baldus S, Bauersachs J et al; ESC/EACTS Scientific Document Group. 2021 ESC/EACTS Guidelines for the management of valvular heart disease. Eur Heart J 2022;43:561–632. [PubMed] [Google Scholar]

2. Lee CY, Sauer JS, Gorea HR, Martellaro AJ, Knight PA.. Comparison of strength, consistency, and speed of COR-KNOT versus manually hand-tied knots in an ex vivo minimally invasive model. Innovations (Phila) 2014;9:111–6; discussion 116. [PubMed] [Google Scholar]

3. Delgado V, Ajmone Marsan N, de Waha S, Bonaros N, Brida M, Burri H. et al.; ESC Scientific Document Group. 2023 ESC Guidelines for the management of endocarditis. Eur Heart J 2023;44:3948–4042. [PubMed] [Google Scholar]

4. Leon MB, Piazza N, Nikolsky E, Blackstone EH, Cutlip DE, Kappetein AP. et al. Standardized endpoint definitions for Transcatheter Aortic Valve Implantation clinical trials: a consensus report from the Valve Academic Research Consortium. J Am Coll Cardiol 2011;57:253–69. [PubMed] [Google Scholar]

5. Harnoss JC, Partecke LI, Heidecke CD, Hübner NO, Kramer A, Assadian O.. Concentration of bacteria passing through puncture holes in surgical gloves. Am J Infect Control 2010;38:154–8. [PubMed] [Google Scholar]

6. Rasmussen M, Mohlin AW, Nilson B.. From contamination to infective endocarditis—a population-based retrospective study of Corynebacterium isolated from blood cultures. Eur J Clin Microbiol Infect Dis 2020;39:113–9. [PMC free article] [PubMed] [Google Scholar]

7. Byrd AL, Belkaid Y, Segre JA.. The human skin microbiome. Nat Rev Microbiol 2018;16:143–55. [PubMed] [Google Scholar]

8. Dhom J, Bloes DA, Peschel A, Hofmann UK.. Bacterial adhesion to suture material in a contaminated wound model: comparison of monofilament, braided, and barbed sutures. J Orthop Res 2017;35:925–33. [PubMed] [Google Scholar]

9. Crawford RJ, Webb HK, Truong VK, Hasan J, Ivanova EP.. Surface topographical factors influencing bacterial attachment. Adv Colloid Interface Sci 2012;179–182:142–9. [PubMed] [Google Scholar]

10. Henry-Stanley MJ, Hess DJ, Barnes AM, Dunny GM, Wells CL.. Bacterial contamination of surgical suture resembles a biofilm. Surg Infect (Larchmt) 2010;11:433–9. [PMC free article] [PubMed] [Google Scholar]

11. Katz S, Izhar M, Mirelman D.. Bacterial adherence to surgical sutures. A possible factor in suture induced infection. Ann Surg 1981;194:35–41. [PMC free article] [PubMed] [Google Scholar]

12. Zheng S, Bawazir M, Dhall A, Kim HE, He L, Heo J. et al. Implication of surface properties, bacterial motility, and hydrodynamic conditions on bacterial surface sensing and their initial adhesion. Front Bioeng Biotechnol 2021;9:643722. [PMC free article] [PubMed] [Google Scholar]

13. Moreillon P, Que YA.. Infective endocarditis. Lancet 2004;363:139–49. [PubMed] [Google Scholar]

14. Beute TJ, Oram MD, Schiller TM, Goehler M, Parker J, Willekes CL. et al. Use of an automated suture fastening device in minimally invasive aortic valve replacement. Ann Thorac Surg 2018;106:58–62. [PubMed] [Google Scholar]

15. Plestis K, Orlov O, Shah VN, Wong J, Thomas M, Aharon A. et al. Facilitating technologies in minimally invasive aortic valve replacement: a propensity score analysis. Interact CardioVasc Thorac Surg 2018;27:202–7. [PubMed] [Google Scholar]

16. Dangas GD, Weitz JI, Giustino G, Makkar R, Mehran R.. Prosthetic heart valve thrombosis. J Am Coll Cardiol 2016;68:2670–89. [PubMed] [Google Scholar]

17. Sánchez J, Lundquist PB, Elgue G, Larsson R, Olsson P.. Measuring the degree of plasma contact activation induced by artificial materials. Thromb Res 2002;105:407–12. [PubMed] [Google Scholar]

18. Li B, Bai S, Yue G, Zhu J, Zhang M, Yang B. et al. Safety and effectiveness outcomes of a novel automated titanium suture fastener device applied for heart valve surgery in an ovine model. Front Cardiovasc Med 2022;9:783208. [PMC free article] [PubMed] [Google Scholar]

19. Sazzad F, Ler A, Kuzemczak M, Ng S, Choong A, Kofidis T.. Automated fastener vs hand-tied knots in heart valve surgery: a systematic review and meta-analysis. Ann Thorac Surg 2021;112:970–80. [PubMed] [Google Scholar]

20. Baciewicz FA Jr. Cor-knot perforation of aortic valve leaflet. Ann Thorac Surg 2018;106:936–7. [PubMed] [Google Scholar]

Articles from European Journal of Cardio-Thoracic Surgery : Official Journal of the European Association for Cardio-thoracic Surgery are provided here courtesy of Oxford University Press

Automated titanium fastener for surgical aortic valve replacement—preventive role for infective endocarditis? (2024)
Top Articles
Latest Posts
Article information

Author: Fr. Dewey Fisher

Last Updated:

Views: 6120

Rating: 4.1 / 5 (62 voted)

Reviews: 85% of readers found this page helpful

Author information

Name: Fr. Dewey Fisher

Birthday: 1993-03-26

Address: 917 Hyun Views, Rogahnmouth, KY 91013-8827

Phone: +5938540192553

Job: Administration Developer

Hobby: Embroidery, Horseback riding, Juggling, Urban exploration, Skiing, Cycling, Handball

Introduction: My name is Fr. Dewey Fisher, I am a powerful, open, faithful, combative, spotless, faithful, fair person who loves writing and wants to share my knowledge and understanding with you.